enviroFlu
enviroFlu

enviroFlu

PAH, oil-in-water using UV fluorescence enviroFlu-HC is the new generation of immersion sensors for measurement of oil-in-water. The used measuring principle of UV fluorescence is much more sensitive than the conventionally used infrared scattering or absorption…  Read more

Measurement technology
  • light source: Xenon flash lamp + filter (254 nm)
  • detector: Photo diode + filter (360 nm)
Measurement principle Fluorescence
Parameter PAH, oil
Measurement range
  • enviroFlu-HC 500
  • PAH: 0...50 ppb, 0...500 ppb
  • Oil: 0...1.5 ppm, 0...15 ppm typical
  • enviroFlu-HC 5000
  • PAH: 0...500 ppb, 0...5000 ppb
  • Oil: 0...15 ppm, 0...150 ppm typical
Detection limit
  • enviroFlu-HC 500 0.3 ppb
  • enviroFlu-HC 5000 0.5 ppb
Measurement accuracy ± 5 % FS
Reproducibility ≤ 0,5 % FS
Turbidity compensation NO
Data logger No
T100 response time ≤ 10 s
Measurement interval ≥ 5 s
Material
  • Housing: Stainless steel (1.4571/1.4404) or titanium (3.7035)
  • Head: POM black with synthetic quartz glass
Dimensions (L x Ø) 311 mm x 68 mm ~ 12.2˝ x 2.6˝
Weight

stainless steel: ~ 2.7 kg

titanium: ~ 1.9 kg

~ 6 lbs

~ 4.2 lbs

Interface
  • digital: RS-232 (TriOS)
  • analog: 4...20 mA, 0...5 V
Power consumption ≤ 3.5 W
Power supply 12...24 VDC (± 10 %)
Required supervision ≤ 0.5 h/month (typical)
Calibration/maintenance interval 24 months
System compatibility Analog Out (0...5 VDC, 4...20 mA)
Warranty
  • 1 year (EU: 2 years)
  • US: 2 years
Installation  
Max. pressure

with SubConn: 30 bar

with fixed cable: 3 bar

in FlowCell: 1 bar, 2...4 L/min

Deepsea version: 600 bar

~ 435 psig

~ 43.5 psig

~ 14.5 psig at 0.5 to 1 gpm

 

Protection type
  • IP68
  • NEMA 6P

Sample temperature

  • +2...+40 °C
  • ~ +36 °F to +104 °F
Ambient temperature
  • -5...+55 °C (2...+40 °C for specified accuracy)
  • ~ +23 °F to +131 °F (~ 35,6 °F to 104 °F for specified accuracy)
Storage temperature
  • -20...+80 °C
  • ~ -4 °F to +176 °F
Inflow velocity
  • 0.1...10 m/s
  • ~ 0.33 fps to 33 fps
Max. immersion depth
  • 30 m with fixed cable ~ 98.43 ft with fixed cable
  • optional: 6000 m deepsea version optional: ~ 19685.04 ft deepsea version

PAH, oil-in-water using UV fluorescence

enviroFlu-HC is the new generation of immersion sensors for measurement of oil-in-water. The used measuring principle of UV fluorescence is much more sensitive than the conventionally used infrared scattering or absorption method. This allows to determine even the slightest traces of PAH's, for example in drinking water and cooling water condensates. Application areas include the petrochemical industry, leakage detection in cooling and wastewater streams as well as environmental monitoring. The devices enable both stationary use in shafts, flows or piping, and mobile use through an optional hand-held measuring instrument. An innovative coating reduces fouling of the optical measuring window and minimizes the maintenance.

  • Without sampling and preparation of test samples
  • Real time sensor
  • Without reagents
  • High sensitivity and selectivity
  • Optical window with nano coating

Drinking water

  • Wastewater
  • Airports
  • Cooling water
  • Desalination plants
  • Refineries
  • Pipeline monitoring
  • Bilge water monitoring
  • Exhaust gas cleaning with approval for ship use
  • according to IMO regulation MEPC.184(59)